Product Code: 1176
The Gemini R is a blackbody source with which radiation pyrometers, commonly referred to as IR
sensors, can be calibrated. With a temperature range of 30 to 550°C, or 700°C, the
temperature of the cavity can be set with 0.1°C resolution. The temperature of the
furnace is set and indicated on a PID controller. An independent indicator is also provided for
higher accuracy with a standard probe, which indicates the actual radiance temperature.
Temperature variation over a 5-minute period, as viewed on the temperature controller, is
±0.2°C and a radiation thermometer directed into the cavity detects
similar variations. The Gemini R’s cavity’s size is 2.5″ Ø x 6.3″ deep (65mm x 160mm). Maximum temperature
differences are ±1°C in the range of 100 to 500°C. The Gemini R is capable
of accepting blackbody fixed points to achieve the highest level of a radiation measurement
possible. Typical fixed-point uncertainties are ±0.25°C. Gallium, indium, tin and
zinc cells are available. Emissivity is greater than 0.995, and a calibration certificate is available for the built-in indicator
and/or supplied probe. The Gemini R can be supplied with RS232 or RS422 communications. Cal Notepad software now comes standard with the Isotech 976 Gemini R. Cal NotePad Software
automatically logs and displays the temperature of an Isotech calibration bath together with the sensor under test. Cal
Notepad can be used with baths (or indicators) without PC interfaces by the user typing in values from the keyboard.
GF Signet U1000 UltraFlow Ultrasonic Flow Sensor
Rosemount 148 Temperature Transmitter
KROHNE ENVIROMAG 2000 Electromagnetic Flow Meter
Mahr MarSurf PS 10 Roughness Tester
Isotech Gold / Platinum Thermocouple Standard
Rosemount 1056 Intelligent Four-Wire Transmitter
Panametrics HygroPro II Moisture Transmitter
Aalborg GFC Mass Flow Controllers
E+E EE310 Humidity & Temperature Transmitter
Bently Nevada Model 9100D Shaker Table
Rosemount 3051S Coplanar Pressure Transmitter
Fluke Calibration Portable Infrared Calibrator
Ashcroft EL Series Bimetal Thermometers
Fluke Calibration 1551A and 1552A “Stik” Thermometers
MTL 4500 Intrinsically Safe Isolators